Status

SCPI Command :

SBUS<*>:TNOS:FRAMe<*>:STATus
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get(serialBus=SerialBus.Default, frame=Frame.Default) SbusTnosFrameState[source]
# SBUS<*>:TNOS:FRAMe<*>:STATus
value: enums.SbusTnosFrameState = driver.sbus.tnos.frame.status.get(serialBus = repcap.SerialBus.Default, frame = repcap.Frame.Default)

Returns the state of the frame.

Parameters:
  • serialBus – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Sbus’)

  • frame – optional repeated capability selector. Default value: Ix1 (settable in the interface ‘Frame’)

Returns:

frame_state: OK | EPRMble | ESFD | EESD | ECRC | ELEN | INComplete

  • OK: The frame has no error, it is valid.

  • EPRMble: Preamble error, the hex value of the preamble field is different from 0x55555555555555 (56 alternating 0’s and 1’s) .

  • ESFD: SFD error, the hex value of the SFD field is different from 0xD5

  • EESD: ESD error, the value of the ESD field does not correspond to the symbol pair ‘ESD, ESDOK’

  • ECRC: CRC error, the value of the FCS field does not match the calculated CRC.

  • ELEN: Length error, the number of bits in the specified frame is higher or lower than expected.

  • INComplete: The frame is incomplete