Status
SCPI Command :
SBUS<*>:I3C:FRAMe<*>:STATus
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(serialBus=SerialBus.Default, frame=Frame.Default) SbusI3cFrameState[source]
# SBUS<*>:I3C:FRAMe<*>:STATus value: enums.SbusI3cFrameState = driver.sbus.i3C.frame.status.get(serialBus = repcap.SerialBus.Default, frame = repcap.Frame.Default)
Returns the overall state of the frame.
- Parameters:
serialBus – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Sbus’)
frame – optional repeated capability selector. Default value: Ix1 (settable in the interface ‘Frame’)
- Returns:
frame_state: OK | ACK | PAR | CRC | LENGth | UNKNown | INComplete
OK: The frame is valid.
ACK: Wrong acknowledgment (ACK) delimiter occurred.
PAR: Wrong parity detected.
CRC: Cyclic redundancy check failed.
LENGth: Wrong length of frame.
UNKNown: Unknown frame type detected.
INComplete: The frame is not completely contained in the acquisition. The acquired part of the frame is usually valid, but cannot always be trusted.