Upper

SCPI Command :

TRIGger:EVENt<*>:LEVel<*>:SLEW:UPPer
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get(evnt=Evnt.Default, lvl=Lvl.Default) float[source]
# TRIGger:EVENt<*>:LEVel<*>:SLEW:UPPer
value: float = driver.trigger.event.level.slew.upper.get(evnt = repcap.Evnt.Default, lvl = repcap.Lvl.Default)

Sets the upper voltage threshold. When the signal crosses this level, the slew rate measurement starts or stops depending on the selected slope.

Parameters:
  • evnt – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Event’)

  • lvl – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Level’)

Returns:

level: -10 to 10

set(level: float, evnt=Evnt.Default, lvl=Lvl.Default) None[source]
# TRIGger:EVENt<*>:LEVel<*>:SLEW:UPPer
driver.trigger.event.level.slew.upper.set(level = 1.0, evnt = repcap.Evnt.Default, lvl = repcap.Lvl.Default)

Sets the upper voltage threshold. When the signal crosses this level, the slew rate measurement starts or stops depending on the selected slope.

Parameters:
  • level – -10 to 10

  • evnt – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Event’)

  • lvl – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Level’)